Sensovation Infos

It is not recommanded to use these functions if you are not sure about the consequences.

Standard Camera Settings

  Camera IP:                            192.168.1.117
  Camera Port:                          19383

Setup Cooler Calibration

  Query Calibration (Standard "5"):     CAL:PID:LIM?
  Boost Calibration (max 20):           CAL:PID:LIM 20
  Store Data:                           SYST:SAVE

Reset Cooler Error Counter

  LogOn:                                serv:cen letmework
  Reset Cooler Error Counter:           SENSor:SYSCheck:REQuired 0

Getter

   Start vacuum activation:(1)          SENSor:VACuum:REACtivate 'Date'
   Query progress:                      SENSor:VACuum:REACtivate?
   Query reactivation stamps:           SENSor:VACuum:DATE[0-19]?
   Query number of reactivations:       SENSor:VACuum:COUNt?
   Reset number of reactivations:(2)    SENSor:VACuum:COUNt 0
  1) 32-bit integer, seconds since 01.01.1970 00:00:00
  2) Requires Logon through SERVice:CENable 'LetMeWork' 