The LumiBIS is an fully automated imaging system for a wide range of samples stained with fluorescent and colorimetric (visible) dyes. The system is equipped with a high resolution and sensitive CCD camera. The LumiBIS features DNR's unique Smart Dark Chamber technology with an automated UV protection mechanism.
Fully Automated System with high resolution
The LumiBIS captures the sharpest possible images with precise sample separation - thanks to its powerful CCD camera with high resolution. Two different models are available, 1.4 and 2 Mpixels. The high speed zoom lens results in enhanced sensitivity, collecting maximum light from the sample, while maintaining sharp, clear images with minimal geometric and light distortion. Featuring a motorized zoom, iris and focus, the LumiBIS is a fully automated system for user comfort and real-time image capturing.
Illumination
The LumiBIS is equipped with a built-in transilluminator with dual UV/WL illumination and 2 interchangeable UV and WL drawers. Automatic Drawer Detection technology indentifies the type of drawer in use and adjusts the software settings accordingly. The motorized 6-position filter wheel is equipped with 3 filters in orange, green and red. A WL epi illumination is installed, a UV epi illlumination can be addded optionally. The optional HeptaLED system provides illumination of 6 different wavelengths at 470nm, 500nm, 525nm, 568nm, 588nm and 621nm.
Applications
The LumiBIS has extensive illumination options allowing researchers to document a wide range of samples. With trans and epi illumination sources, six-position filter wheel and an unique LEDs System (HeptaLED System), you can use this system for almost any fluorescent or colorimetric application.
- Detection and quantitation of nucleic acid and protein stains and labels including: Ethidium Bromide, SYBR™ Green, GelStar™, Fluorescein, Coomassie Brilliant Blue, SYPRO™ Orange, SYPRO™ Red, Silver Stain, SYPRO™ Ruby, Deep Purple™ and many more
- Detection of advanced fluorescent molecules, including Cy2™, Cy3™, Pro-Q™ Diamond, Deep Purple™, Rhodamine Red™ and more
- Other Biological Methods: TLC plates, Dot and slot blots, Autoradiographs, Microplate, Macroarrays, Membranes and more
Ease-of-Use
Easy to set-up, with an user-friendly interface, and one-click results, the LumiBIS ensures greater efficiency and productivity in the laboratory. Fully controlled by the PC, LumiBIS USB 2.0 interface connects easily to any laptop or PC.
Software
The LumiBIS will be delivered standardly with two software packages, GelCapture for imaging acquisition and TL-100 for image analysis.
GelCapture offers complete, automatic control of configuring, capturing, and saving sample images for analysis. GelCapture offers all important image settings adjustable as illumination, focus, brightness, binning and many more.
The TL100 software from the TotalLab range enables fast and consistent analysis of your gel images with high levels of automation and accuracy. You can perform a complete analysis with a single click and then refine results afterwards, or step through the different analysis stages giving you complete control at each stage. Automated lane creation, band detection and background subtraction are all performed without any user intervention for the utmost in analysis consistency. Molecular weight can be calculated for your bands from the standard marker lane you included on the gel. Classical commercial standards are available in the software or you can create and save your own. The TL100 software performs precise band edge detection and subtraction of background material which is vital for correct quantitation or normalisation of band material. There are additional tools to adjust band edges and account for band "grimacing" on awkward gels.
Also included with the TL100 software are modules for:
- Array, dot / slot blot analysis
- Colony counting
- A toolbox for general image analysis
Specifications
Camera and Optics CCD resolution:1360(H) x 1024(V); 1.4 Mpixels or 1600(H) x 1200(V); 2 Mpixels Gradation: 16 bit file format Dynamic range: 3.4 orders of magnitude Binning: 1x1, 2x2, 3x3, 4x4, 5x5, 6x6, 7x7, 8x8 Super bright lens: F/1.2, 8-48mm Illumination and Darkroom Dark chamber: DNR's Smart Dark Chamber Emission filters: 6 position filter wheel with 3 filters included as a standard Illumination modes: Trans-UV, Trans-WL, Epi-WL optional: Epi-UV Excitation sources: White Light source, 312nm optional: 254nm, 365nm optional: dual UV source optional: Monochromatic LEDs system (Hepta LED System) Software Image capture: GelCapture, lifetime free upgrade Image analysis: TL-100 for 1D analysis of Nonlinear Dynamics optional: GelQuant, TL-120, TL-120 DM
The MiniBIS and MiniBIS Pro products are available in Germany, The Netherlands, Belgium, Luxemburg, Italy, India, France, Switzerland, UK, Japan and USA.