Radiation Protection
Home
Contact
Products
Applications
LB150 DR Alpha-Beta Particulate Monitor

Application
The LB150 Particulate Monitor products family can handle the entire range of room and stack air monitoring. The LB150R is intended for monitoring soleley alpha-activity, while the LB150D-R is for simultaneous alpha and beta monitoring. Typical locations include:

  • nuclear fuel cycle facilities
  • radioactive waste processing and storing sites
  • hot cells
  • nuclear decommissioning
  • any site requiring monitoring of alpha- or beta-emitting particulates

Measuring Principles
The LB150 Particulate Monitors are fixed filter, on-line monitors. They are sampling the airborne radioactivity, and measureme and analyze the activity concentrations.

Sampling:
The high air flow rate, 40m³/h (28 cfm), allows  representative sampling of room air or isokinetic sampling of air from a stack.

Dust Collection and Choice of Detector: The air is drawn through a 200 mm (8") diameter, S-class glass fiber filter, which is viewed by a thin-window, large-area, proportional counter located directly above the filter. For the LB150R Particulate Monitor, two stacked counter tubes are used; for the LB150DR, a triple counter tube stack is used.

The air-tight dust collection unit may be made of either stainless steel or teflon-coated stainless steel.



Technical Data