Radiation Detector: Large area scintillator (ZnS:Ag) Light Detection: PMT Measuring Mode: simultaneous and seperate alpha/beta measurement Sensitive Area: approx. 345 cm² Window Thickness: 6 µm aluminized plastic (0.4 mg/cm²) Grid transmission: approx. 80%
Efficiency (relative to activity of a reference source with 100 cm² area): C-14 (Beta channel): 10% Cl-36 (Beta channel): 45% Tc-99m (Beta channel): 10% Am-241 (Alpha channel): 22% Background: approx. 0.1 cps (Alpha channel) approx. 15 cps (Beta channel) Measuring Range: 0 to 5000 cps (Alpha channel) 0 to 50000 cps (Beta channel)
Temperature range operation: -20°C to +40°C Temperature range storage: -40°C to +60°C
Humidity: 0% to 95% (without non condensation) External pressure operation: 500 to 1300 hPa External pressure storage: 100 to 1300 hPa
Protection: IP 54
Operational life time with rechargeable batteries (without illumination): > 25 h Operational life time with batteries (without illumination): > 65 h
Weight: 1850 g with batteries Selectable calibration factors: more than 50 nuclides Averaging method: selectable (ratemeter, scaler timer) Measuring-/Averaging time: selectable Data storage: > 1000 measuring values Data output: serial output (RS232)
|