Basic Unit LB 1230
Display: High-contrast dot-matrix with 32 x 84 pixels Background illumination can be switched off 4-digits numerical values with autoranging
Control: Membrane keypad comprising 1 on/off button and 4 soft keys
Detector connection: 8-pin connector socket Fischer 04, spiral cable LB 75576 as standard
Data output (Option): FSMA connector, via beam wave guide and beam wave guide interface LB 75306 with D 25 connector
Temperature range: -15°C to 50°C
Dimensions: 145 mm (H) x 170 mm (W) x 45 mm (D)
Weight: approx. 800 g (with batteries)
Power supply: 3 x IEC-R14 (baby cell) or 3 x rechargeable cells Varta NiCd #5014
Maximum operating: with R14 > 150 hours
Probe Adapter LB 7637 for Contamination Probes
Dimensions: 160 mm (h, with handle), 140 mm (w), 240 mm (D)
Weight: 1,5 kg including detector
Wall Bracket LB 1250
Low voltage supply: + 5 V via mains adapter LB 7619
Dimensions: 68 mm (h) x 245 mm (W) x 77 mm (D)
Transport Case Dimensions: 180 mm (h) x 450 mm (W) x 350 mm (D)
Contamination Probes LB1231, LB1232 and LB1233
LB1231 Type of radiation: Beta and Gamma Counter tube: LB 6357 Counting gas: Xenon, sealed Window area: 120 x 190 mm² Window thickness: 5 mg/cm² Temperature range: -15°C to 50°C
LB1232 Type of radiation: Alpha and Beta Counter tube: LB 6358 G Counting gas: Butane, for manual refilling Window area: 120 x 190 mm² Window thickness: 0,4 mg/cm² Temperature range: 5°C to 30°C
LB1233 Type of radiation: Alpha and Beta Counter tube: LB 6359 Counting gas: P-10, flow-through Window area: 120 x 190 mm² Window thickness: 0,4 mg/cm² Temperature range: -15°C to 30°C
Alpha-Beta Activity Measurement Probe LB1238 Counter tube: End-window proportional counter tube Window: 28 mm Ø, 2 mg/cm² Efficiency for point sources: Sr-90 46%, Am-241 18% Temperature range: -10°C to 50°C Dimensions: 50 mm Ø x 275 mm Weight: 0,46 kg
Tritium Probe LB1239 Counter tube: open proportional counter tube Counting gas: P10, flow-through Entrance window: 150 mm x 15 mm External dimensions: 226 mm x 86 mm x 34 mm without handle Sensitivity: approximately 1 cps per 2.2 Bq Temperature range: -10°C to 50°C Weight: 0,90 kg
|
|
|