|
|
 |
Technical Data
|
 |
 |
The personal contamination monitors LB 145 and LB 146 comply with or exceed the requirements of IEC 1098. This is true in particular for the detector performance, the manufacturers’ calibration procedures and the determination of counting efficiencies.
LB145 probes The counters with integrated preamplifier, Xenon as counting gas and a window of titanium foil with a density of 5 mg/cm².
LB146 probes The counter tubes are large area proportional counters with a window foil of 0.4 mg/cm² aluminated plastic and an integrated preamplifier, operated in the gas flow mode. As counting gas, the non-explosive and inflammable mixture of Ar/CO2 with 18% Co2 share is recommended. However, the monitor LB 146 can also be operated with the more common P-10 gas, a mixture of 90% Ar and 10 % methane. Thanks to the electronically controlled gas flow regulation, the gas consumption is reduced to about 1 to 2 litres per hour.
LCD colour display The VGA compatible LCD colour display measures 195 mm x 145 mm. An external video monitor is optional.
Dimensions and background count rates
Detectors LB145 Hand Probes LB6370: Window dimensions: 150 x 230 mm Window Area: 345 cm² Protective Grid Permeability: 78% Background Countrate: < 25 ips
Foot Probes LB6372: Window dimensions: 150 x 350 mm Window Area: 525 cm² Protective Grid Permeability: 55% Background Countrate: < 40 ips
Clothing Probe LB6371: Window dimensions: 150 x 70 mm Window Area: 105 cm² Protective Grid Permeability: 78% Background Countrate: < 18 ips Efficiency (%) and Detection Limit (Bq/cm²) for frequently used nuclides
C-14: LB 6370: 4,0 and 1,3 LB 6372: 1,6 and 4,2 LB 6371: 3,2 and 2,3
CI-36: LB 6370: 24,5 and 0,2 LB 6372: 17,4 and 0,4 LB 6371: 21,0 and 0,3
SR-90: LB 6370: 30,5 and 0,2 LB 6372: 18,5 and 0,4 LB 6371: 21,7 and 0,3
TC-99: LB 6370: 11,1 and 0,5 LB 6372: 6,3 and 1,1 LB 6371: 9,3 and 0,8
I-125: LB 6370: 3,1 and 1,5 LB 6372: 2,2 and 2,5 LB 6371: 2,8 and 2,1
CS-137: LB 6370: 26,8 and 0,2 LB 6372: 16,4 and 0,4 LB 6371: 20,2 and 0,4
Pm-147: LB 6370: 9,5 and 0,5 LB 6372: 8,2 and 0,8 LB 6371: 6,6 and 1,1
TI-204: LB 6370: 20,3 and 0,3 LB 6372: 14,5 and 0,5 LB 6371: 16,4 and 0,4
Am-241: LB 6370: 12,1 and 0,4 LB 6372: 2,3 and 2,9 LB 6371: 0,0 and 0,7
Detectors LB146 Hand Probes LB6373: Window dimensions: 150 x 230 mm Window Area: 345 cm² Protective Grid Permeability: 78% Background Count rate: Alpha 0,05 ips / Beta <12 ips
Foot Probes LB6375: Window dimensions: 150 x 350 mm Window Area: 525 cm² Protective Grid Permeability: 55% Background Count rate: Alpha 0,1 ips / Beta <20 ips
Clothing Probe LB6374: Window dimensions: 150 x 70 mm Window Area: 105 cm² Protective Grid Permeability: 78% Background Count rate: Alpha 0,03 ips / Beta <5 ips
Efficiency (%) and Detection Limit (Bq/cm²) for frequently used nuclides
C-14: LB 6373: 17,0 and 0,25 LB 6375: 8,2 and 0,67 LB 6374: 12,5 and 0,43
CI-36: LB 6373: 38,5 and 0,11 LB 6375: 25,0 and 0,22 LB 6374: 28,6 and 0,2
Co-60: LB 6373: 25,0 and 0,17 LB 6375: 15,6 and 0,35 LB 6374: 18,5 and 0,3
SR-90: LB 6373: 37,0 and 0,12 LB 6375: 25,0 and 0,22 LB 6374: 28,6 and 0,2
TC 99: LB 6373: 25,0 and 0,17 LB 6375: 16,0 and 0,35 LB 6374: 20,0 and 0,27
CS-137: LB 6373: 35,7 and 0,12 LB 6375: 23,8 and 0,23 LB 6374: 27,0 and 0,2
TI-204: LB 6373: 34,5 and 0,2 LB 6375: 22,7 and 0,24 LB 6374: 25,0 and 0,2
Po-210: LB 6373: 25,6 and 0,02 LB 6375: 14,5 and 0,046 LB 6374: 19,0 and 0,08
Am-241: LB 6373: 23,2 and 0,024 LB 6375: 13,9 and 0,05 LB 6374: 17,2 and 0,09
LB145/LB146 Dimensions: Foot Pedestal: 516 mm (W) x 874 mm (D) Hand Desk: 700 mm (W) x 410 mm (D) Complete Monitor: 700 mm (W) x 874 mm (D) x 1283 mm (H)
Weight: 81 kg
Power Consumption: 150 mA
|
|
|
|
 |
|
|