Radiation Detectors: Large area scintillator (ZnS:Ag) Light Detection: PMT Measuring Mode: simultaneous and separate alpha/beta measurement Measuring Range: 0 to 50000 cps (dead time <10%) Temperature range: -10°C to +40°C Humidity: 0% to 95% (no condensing)
Hand Detectors Sensitive Area: 345 cm² Window Thickness: 6 µm aluminized plastic (0.4 mg/cm²) Efficiency: (rel. to the activity of a 100 cm² source)
14C (Beta channel): > 6% 99mTc (Beta-channel): > 5% 36Cl (Beta channel): » 30% 241Am (Alpha channel): »15%
Background (Alpha channel): approximately 0.1 cps Background (Beta channel): approximately 15 cps
Foot Detectors Sensitive Area: 555 cm² Window Thickness: 6 µm aluminized plastic (0.4 mg/cm²) Efficiency: (rel. to the activity of a 100 cm² source)
14C (Beta channel): > 6% 99mTc (Beta-channel): > 5% 36Cl (Beta channel): » 30% 241Am (Alpha channel): »8%
Background (Alpha channel): approximately 0.2 cps Background (Beta channel): approximately 40 cps
|