Our portable contamination monitor LB 124 SCINT offers you the following features:

  • Simultaneous separate measurement of α- and β-γ radiation
  • Calibration factors for over 50 nuclides
  • Calibration optionally according to ISO 7503-1 or with reference to an activity on 100 cm2
  • Protective grid with high transparency
  • Adjustable alarm thresholds
  • Acoustic alarm
  • High sensitivity and uniform response
  • No counting gas required
  • Lightweight, handy and extremely robust device
  • Application over a wide temperature range
  • portable battery-operated meter
  • Easily decontaminated device surface

LB 124 SCINT

Details

The contamination monitor LB 124 SCINT is a portable battery-operated measuring device. It consists of a display unit with microprocessor electronics, signal processing electronics and a ZnS scintillation detector with photomultiplier with an active measuring area of 170 cm2. Due to a sophisticated reflector geometry, the device has an extremely uniform response over the entire surface. The device has guides on its underside. These allow the attachment of an additional protective grid or the use of an accessory drawer for activity measurements of small samples.

Highlights

Simultaneous separate measurement of α- and β-γ radiation

High sensitivity and uniform response

Light, handy and extremely rugged device

Technical Details

Radiation detectorZnS(Ag) scintillator
Measurement modes

Simultaneous α- and β-γ-measurement;
ratemeter, scaler-timer-mode,
 

clearance measurement, survey mode

Dimensions entrance window118 mm x 145 mm
Sensitive area170 cm²
Material entrance window2 x 3 µm metallized plastic (0.4 mg/cm²)
Protective grid80 % transmission
Typical background (α-channel)                     approx. 0.1 cps
Typical background (β-γ-channel)approx. 10 cps
Typical efficiencies (according
to ISO 7503-1) 239Pu 

approx. 41 %

Typical efficiencies (according to ISO 7503-1) 241Amapprox. 44 %
Typical efficiencies (according to ISO 7503-1) 14Capprox. 29 %
Typical efficiencies (according to ISO 7503-1) 36Cl approx. 69 %
Typical efficiencies (according to ISO 7503-1) 60Coapprox. 58 %
Typical efficiencies (according to ISO 7503-1) 137Csapprox. 71 %
Gamma sensitivity at 1 µSv/h 137Cs (α-channel)Not detectable
Gamma sensitivity at 1 µSv/h 137Cs (β-γ-channel)< 100 cps
Spillover α- in β-γ-channel< 20 %
Spillover β-γ- in α-channel< 2 x 10-5
Measuring range (α-channel)0 to 5000 cps
Measuring range (β-γ-channel)0 to 50000 cps

 

Temperature range- 20 °C to + 40 °C (operation)
Relative humidity

0 % to 80 %, no condensation

External pressure500 hPa to 1300 hPa (operation)
Protection classIP 53 (according to IEC 60529)

 

Display

Monochrome LCD 192 x 64 pixels 

Electro-luminescence-illumination

InterfacesRS 232, headphone connection
Power supply

3 x batteries type "C", Baby or NiMH batteries,

Rechargeable by plug type power supply or
alternatively in the wall mounting bracket

Max. operating time (without illumination)

>50 h with alkali batteries 7.8 Ah

>25 h with NiMH rechargeable batteries 4.5 Ah

Data memory1000 measured values with date and time
AlarmAcoustic with adjustable alarm thresholds

 

Dimensions (L x W x H in mm)

240 x 140 x 110
Weight (with batteries)approx. 1300 g

 

Lorem ipsum dolor sit amet, consetetur sadipscing elitr, sed diam nonumy eirmod tempor invidunt ut labore et dolore magna aliquyam erat, sed diam voluptua. At vero eos et accusam et justo duo dolores et ea rebum. Stet clita kasd gubergren, no sea takimata sanctus est Lorem ipsum dolor sit amet. Lorem ipsum dolor sit amet, consetetur sadipscing elitr, sed diam nonumy eirmod tempor invidunt ut labore et dolore magna aliquyam erat, sed diam voluptua. At vero eos et accusam et justo duo dolores et ea rebum. Stet clita kasd gubergren, no sea takimata sanctus est Lorem ipsum dolor sit amet.