Download nowOur custom-made rod sources have the radioactive material distributed along the entire source length tailored upon the application, vessel geometry, and detector. Compared to a point source, the individual activity distribution improves the linearity of the measured count rate at the detector to the level in the vessel for continuous level measurements and ensures the highest accuracy with the simplest calibration.
With our unique manufacturing technology, the radioactive material is distributed along an inner guiding rod within a tightly sealed outer capsule. In a rod source / point detector arrangement, the radioactive material in the rod source is higher concentrated in the lower part of the rod, which intensifies the activity in this area. This is advantageous because due to the increasing distance to the detector, which is usually mounted at the top, the travel distance of the gamma radiation increases, as well as the radiation must pass through the vessel walls at a smaller angle. This increases the effective wall thickness and thus the absorption. A higher activity in the lower part of the rod source compensates for this effect.
AN ARRANGEMENT WITH A ROD SOURCE SHOULD BE CHOSEN,