Highlights
General
- Monitors radioactive gases such as Krypton-85, Argon-41, Xenon-133, Carbon-11 and Fluorine-18 in rooms
- 1 to 4 large area proportional counter tubes, xenon sealed
- Measurement of beta- and beta+ (positrons) nuclides
Features
- Integrated preamplifiers and high voltage generators
- Measuring chamber with 11.45 l measuring volume
Product Overview

BAI 9109-4
- The Beta-Gasmonitor BAI 9109-4 provides the opportunity to monitor radioactive gases such as Krypton-85, Argon-41, Xenon-133, Carbon-11 and Fluorine-18 in rooms and discharged air with low minimal detectable activities.
- The sample air being measured will be directed via a prefilter by means of a pump disturbing particles within the sample air will be removed and the air will be directed to the measuring chamber.
- The measuring chamber has a volume of 11.45 l and is equipped with a lead shield all sides with 2 cm thickness in 4π geometry to minimize the background as much as possible.
Technical Data
Measuring Chamber BAI 9109-4 | |
| Construction material | PVC |
| Volume | 11.45 l |
| Pump | max. nominal flow rate through the vessel 15 m³/h |
| Lead Shield | |
| Type | 96 % Pb + 4 % Sb |
| Geometry | 4π – all sided with 2 cm thickness |
| Weight | approx. 185 kg (lead only) |
| Detector | |
| Type | sealed xenon large area proportional counter tubes LB 6350-3 |
| Sensitive detector area | 230 cm² |
| Measuring range | 1 kBq/m³ to 100 MBq/m³ |
| Ambient Conditions | |
| Temperature | 0°C to +40°C |
| Humidity | 10 % to 95 % (non-condensing) |